Product Details: pH-6001 Plastic-Cased pH Electrode
pH Electrode Measurement Principle
The pH-6001 plastic-cased pH electrode operates on the principle of electrochemistry, specifically leveraging the galvanic cell concept. A galvanic cell is a system designed to convert chemical energy into electrical energy. The voltage generated by this cell is known as the Electromotive Force (EMF). The EMF consists of two half-cells: one is referred to as the measuring half-cell, whose potential is related to the specific ion activity; the other is the reference half-cell, commonly known as the reference electrode, which is typically in communication with the measured solution and connected to the measuring instrument. The potential difference produced by the internal galvanic cell reaction is transmitted to the pH controller, which then uses a corresponding algorithm to convert and display the pH value.
Technical Specifications of pH-6001 Plastic-Cased pH Electrode
German Technology: | Utilizes large-section Teflon diaphragm. |
Stable Liquid Junction: | Ensures reliable performance. |
Response Sensitivity: | Achieves 95% of the measurement value within 5 seconds. |
Reference System: | Filled with polymer, providing strong anti-interference ability, resistance to contamination and clogging, and long service life. |
Measurement Range: | (2~12) pH |
Temperature Range: | (0~80) °C |
Pressure Resistance: | 0.4 MPa |
Temperature Compensation Types: | NTC10K, PT100, PT1000 |
Reference Type: | Ag/AgCl |
Salt Bridge: | Ring-shaped Teflon salt bridge |
Installation Thread: | Upper and lower 3/4NPT pipe thread |
Case Material: | ABS |
Application Range: | Water treatment, waste gas treatment, aquaculture, dosing equipment supporting |
Electrode Slope
The theoretical slope of a glass electrode at 25°C is 59.16 mV, meaning that for every change of one pH unit in the solution, the glass electrode produces a potential change of 59.16 mV. However, in practice, no glass electrode can reach 100% of this theoretical value. Typically, the electrode slope is above 95% of the theoretical value (percentage slope). Additionally, the mV difference per pH unit varies with temperature.
Temperature Correction Formula:
ΔE = 59.16 * [(273 + T) / 298] * ΔpH
pH-6001 Plastic-Cased pH Electrode Usage Precautions
Electrode Lead: The electrode lead is a special shielded cable. Customers are prohibited from cutting or splicing the lead without authorization. We will not be responsible for any consequences arising from such actions.
Immersion Environment: Avoid long-term immersion in distilled water or protein solutions, and prevent contact with silicone grease.
Maintenance: For electrodes that have been in use for an extended period, the glass membrane may become transparent or covered with deposits. In such cases, clean the electrode with dilute hydrochloric acid and rinse with water.
Replacement: If the electrode cannot be calibrated or perform normal measurements even after maintenance and servicing, it indicates that the electrode cannot recover its responsiveness. Please replace the electrode.
Water Resistance: The pH electrode cable is not waterproof. Avoid contact between the electrode cable and water.